MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/14992027.2019.1671616

Медицинская библиотека · International journal of audiology · 2019

The Oticon Medical Neuro Zti cochlear implant and the Neuro 2 sound processor: multicentric evaluation of outcomes in adults and children.

Franco-Vidal Valérie, Parietti-Winkler Cécile, Guevara Nicolas, Truy Eric, Loundon Natalie, Bailleux Sonanda, Ardoint Marine, Saaï Sonia, Hoen Michel, Laplante-Lévesque Ariane, Mosnier Isabelle, Bordure Philippe, Vincent Christophe

Оценочное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Objective

This study evaluated the outcomes of the Oticon Medical Neuro Zti cochlear implant and the Neuro 2 sound processor.

Design

Neuro One users were upgraded to Neuro 2. Monosyllabic word identification was evaluated in adults with Neuro One after ≥5 months, with Neuro 2 at upgrade, and with Neuro 2 after 3 months. Self-reported listening ability, satisfaction, and usability were measured in adults and children.Study sample: Participants were 44 adults and 26 children.

Results

Speech identification scores in quiet and noise were 58% and 45% with Neuro One and 67% and 55% with Neuro 2 after 3 months, respectively. Hearing impairment duration and number of active electrodes significantly predicted speech identification in noise with Neuro 2. Significantly higher questionnaire ratings were obtained for Neuro 2 than Neuro One regarding listening ability in complex listening situations, comfort and music, as well as nine aspects of satisfaction and usability.

Conclusion

This study demonstrates the clinical superiority of the Neuro 2 sound processor over Neuro One in terms of speech identification in quiet and in noise and reported patient benefit and satisfaction. Given the study design, sources of improvement may include factors unrelated to the sound processor itself.

Для цитирования: Franco-Vidal Valérie, Parietti-Winkler Cécile, Guevara Nicolas, Truy Eric, Loundon Natalie, Bailleux Sonanda, Ardoint Marine, Saaï Sonia, Hoen Michel, Laplante-Lévesque Ariane, Mosnier Isabelle, Bordure Philippe, Vincent Christophe. The Oticon Medical Neuro Zti cochlear implant and the Neuro 2 sound processor: multicentric evaluation of outcomes in adults and children.. International journal of audiology. 2019. DOI: 10.1080/14992027.2019.1671616.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.