Measurement of Cochlear Implant Electrode Position From Intraoperative Post-insertion Skull Radiographs: A Validation Study.

Svrakic Maja, Friedmann David R, Berman Phillip M, Davis Adam J, Roland J Thomas, Svirsky Mario A

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To validate a method of measuring angular depth of insertion (aDOI) as well as positional depth of each electrode contact in a cochlear implant by using intraoperative postinsertion skull radiographs.

STUDY

DESIGN

Retrospective review.

SETTING

Tertiary referral center.

PATIENTS: Intraoperative postinsertion radiographs obtained from 18 cochlear implant recipients were chosen for analysis. One high-resolution computer tomography scan of the head with the electrode in place was also analyzed. One cadaveric temporal bone with an inserted electrode provided additional data for analysis.

INTERVENTION: aDOI and position of each electrode contact were measured from the radiographs using readily available software. High-resolution computer tomography imaging of the cochlea and electrode were reconstructed in three dimensions and used to simulate head rotation during intraoperative radiographs. The cadaveric temporal bone was imaged by x-ray at various acquisition angles.

MAIN OUTCOME MEASURES: We evaluated the error introduced in measuring aDOI by assessing intra- and inter-rater variability. We also evaluated the error introduced by x-ray acquisition at nonstandardized angles by analyzing the three-dimensional construct and the cadaveric temporal bone.

RESULTS

The concordance correlation coefficients for intrarater (0.991) and inter-rater (0.996) variability in aDOI measurement were excellent. The error introduced by nonstandardized x-ray acquisition angles was only -12.5 degrees to +15.8 degrees even at the limits of clinically relevant head rotation.

CONCLUSIONS

The intraoperative postinsertion radiograph is sufficient for estimating positional depth of electrode contacts and the aDOI. This measure is robust in the face of nonstandardized x-ray acquisition angles, and shows good intra- and inter-rater variability.

Для цитирования: Svrakic Maja, Friedmann David R, Berman Phillip M, Davis Adam J, Roland J Thomas, Svirsky Mario A. Measurement of Cochlear Implant Electrode Position From Intraoperative Post-insertion Skull Radiographs: A Validation Study.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2015. DOI: 10.1097/mao.0000000000000852.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.