K-Y™ jelly inhibits increase in endotracheal tube cuff pressure during nitrous oxide exposure in vitro.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Koyama Yukihide, Oshika Hiroyuki, Nishioka Hiroko, Kamoshida Naoko, Tanaka Sousuke, Inagawa Gaku, Andoh Tomio
Аннотация
BACKGROUND: The increase in endotracheal tube cuff pressure due to nitrous oxide diffusion is a well-known risk during general anesthesia using nitrous oxide. We hypothesized that lubricating endotracheal tube cuffs with K-Y™ Jelly might inhibit the increase in cuff pressure that occurs during exposure to nitrous oxide.
METHODS: We used two types of endotracheal tube cuffs: one made from ultrathin polyurethane (PU) and another made from conventional polyvinyl chloride (PVC). Using a pediatric trachea model, which consisted of an acrylic cylinder with an internal diameter of 12 mm, we measured changes in the cuff pressure during nitrous oxide exposure in size 5.0-mm internal diameter endotracheal tubes with each type of cuff, with and without lubrication with K-Y™ Jelly.
RESULTS: During nitrous oxide exposure, the increase in cuff pressure was significantly lower in the lubricated cuffs than in the non-lubricated cuffs in both types of cuffs (PVC, P < 0.0001; PU, P < 0.0001). However, the cuff compliance in the trachea model was unaffected by lubrication in both types of cuffs.
CONCLUSIONS: Lubrication of endotracheal tube cuffs with K-Y™ Jelly may effectively delay the increase in cuff pressure that occurs during general anesthesia using nitrous oxide.
METHODS: We used two types of endotracheal tube cuffs: one made from ultrathin polyurethane (PU) and another made from conventional polyvinyl chloride (PVC). Using a pediatric trachea model, which consisted of an acrylic cylinder with an internal diameter of 12 mm, we measured changes in the cuff pressure during nitrous oxide exposure in size 5.0-mm internal diameter endotracheal tubes with each type of cuff, with and without lubrication with K-Y™ Jelly.
RESULTS: During nitrous oxide exposure, the increase in cuff pressure was significantly lower in the lubricated cuffs than in the non-lubricated cuffs in both types of cuffs (PVC, P < 0.0001; PU, P < 0.0001). However, the cuff compliance in the trachea model was unaffected by lubrication in both types of cuffs.
CONCLUSIONS: Lubrication of endotracheal tube cuffs with K-Y™ Jelly may effectively delay the increase in cuff pressure that occurs during general anesthesia using nitrous oxide.