DOI 10.1097/00003446-199910000-00001
Objective calibration of bone conductors using otoacoustic emissions.
Аннотация
OBJECTIVE
To demonstrate a technique for objectively calibrating bone conductors on an individual basis using distortion product otoacoustic emissions (DPOAEs).
DESIGN
Individual calibrations were obtained using DPOAEs recorded from a single ear of 21 normally hearing adults. Validity and robustness of the technique were investigated through subjective phase cancellation measurements and sensitivity analysis.
RESULTS
Calibrations obtained using the DPOAE method were well supported by phase cancellation results. Intersession repeatability was good, and manipulation of the DPOAE data showed that the calculated calibration is relatively insensitive to small variations of emission magnitude. Bilateral stimulation through bone conduction did not display an apparent effect on emission magnitude in a single individual.
CONCLUSION
Bone conductors can be accurately calibrated on an individual basis with good repeatability using DPOAEs. The technique is robust and offers an objective, noninvasive calibration method for research and specialized clinical applications. No training and only passive cooperation are required, making the procedure ideal for special groups such as children. A number of limitations will reduce the clinical utility of this technique. Important audiometric frequencies below 1 kHz cannot be tested because of noise, because individuals with significant hearing loss are unlikely to produce sufficient DPOAEs, and because commercial bone conductors typically have poor high-frequency response above 4 kHz.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.