DOI 10.1080/14992027.2022.2034059

Медицинская библиотека · International journal of audiology · 2022

Extended high-frequency bone conduction audiometry Calibration of bone conductor transducers in the conventional and extended high-frequency range.

Rhebergen Koenraad S

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To monitor ototoxicity, air conduction (AC) extended high frequency (EHF) thresholds can be measured up to 16 kHz. However, conductive hearing loss might influence these results. This is unfortunate because the EHF thresholds are important to follow the impact of ototoxic medication during therapy. Therefore a suitable bone conduction (BC) transducer and norm values for EHF BC measurements are needed.

DESIGN

In this study three different BC transducers were used: the B71 (Radioear), the KH70 (Präcitronic), and the KLH96 (Westra). Hearing thresholds were measured from 0.125 to 16 kHz using AC transducers (Telephonics TDH39, Sennheiser HDA200), and BC thresholds from 0.25 to 8 kHz with the B71, and from 0.25 to 16 kHz with the KLH96 and KH70.

STUDY SAMPLE: 60 ears of 30 normal hearing subjects were measured.

RESULTS

The KLH96 showed the highest output for the high frequencies, and distortion measurements were similar to the KH70. The results show that EHF measurements are possible using the KLH96 and KH70 bone conductors.

CONCLUSION

EHF BC measurements are reliable when using the KLH96 and KH70 bone conductors. The extended force sensitivity of the used artificial mastoid should be determined for a proper EHF BC calibration.

Для цитирования: Rhebergen Koenraad S. Extended high-frequency bone conduction audiometry Calibration of bone conductor transducers in the conventional and extended high-frequency range.. International journal of audiology. 2022. DOI: 10.1080/14992027.2022.2034059.