m+Medelevator.RU
← Поиск публикаций
Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons · 1996

Distraction osteogenesis in maxillofacial surgery using internal devices: review of five cases.

Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Chin M, Toth B A

Аннотация

PURPOSE: The purpose of this report is to show the feasibility and potential advantages of using internal devices for distraction osteogenesis in the management of maxillofacial skeletal deficiencies.

PATIENTS AND METHODS: Distraction osteogenesis was used to correct a variety of maxillofacial skeletal deformities in five patients. One patient underwent bilateral Le Fort III advancement aided by distraction, three patients underwent mandibular ramus lengthening, and one patient underwent segmental alveolar reconstruction by distraction. The devices were activated by either a transcutaneous or transmucosal pin. After achievement of the desired skeletal transport, the activating pins were disengaged and removed from the distraction device. This allowed the distraction device to remain submerged and to stabilize the site of the consolidating bone.

RESULTS: All patients achieved lengthening of their jaws. However, premature consolidation was noted in two patients, and one patient had significant relapse.

CONCLUSIONS: Development of internal distraction devices is important to address the limitations of currently available biphasic systems. Potential benefits of internal devices include 1) elimination of skin scarring caused by translation of transcutaneous fixation pins, 2) improved patient compliance during the fixation or consolidation phase because there is no external component, and 3) improved stability of the attachment of the device to the bone.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты