BACKGROUND

Janus kinase inhibitors are antirheumatic immunosuppressive drugs that target intracellular Janus kinases (JAKs). Baricitinib is a selective and reversible orally administered JAK1/JAK2 inhibitor approved for treating rheumatoid arthritis, atopic dermatitis, and alopecia areata in adult patients. Expanded access to baricitinib has been approved for treating pediatric patients affected by rare Mendelian autoinflammatory diseases with type I interferon-mediated damage. Knowledge of the pharmacokinetic properties and target plasma levels of baricitinib in pediatric patients is limited. In this study, a novel LC-MS/MS method for measuring baricitinib in plasma, validated according to the ICH M10 guidelines, is presented.

METHODS

Sample preparation was performed by adding 10 µL of IS working solution (150 ng/mL) and 200 µL of MeOH to each plasma sample. Chromatographic separation was conducted using a Thermo Scientific Accucore Polar Premium column (50 mm × 2.1 mm, i.d. 2.6 m). This method was applied to 7 real anonymous plasma samples obtained from pediatric patients treated with baricitinib at IRCCS Istituto Giannina Gaslini (Genoa, Italy). Patients of both sexes had a median age of 14 years (range, 10-17 years).

RESULTS

The LC-MS/MS method resulted linear over wide concentration ranges (1.024-100 ng/mL) and was accurate and reproducible in the absence of matrix effects, allowing for robust, specific, and rapid quantification of baricitinib from a low amount of plasma (50 µL). The plasma concentration of baricitinib in the samples of the patients, expressed as mean ± SD, was 11.25 ± 10.86 ng/mL.

CONCLUSIONS

This novel LC-MS/MS method is suitable for the therapeutic drug monitoring of baricitinib and can help guide therapy optimization in pediatric patients.

Для цитирования: Cafaro Alessia, Baiardi Giammarco, Pigliasco Federica, Barco Sebastiano, Mattioli Francesca, Volpi Stefano, Caorsi Roberta, Gattorno Marco, Cangemi Giuliana. A Novel LC-MS/MS Method for Therapeutic Drug Monitoring of Baricitinib in Plasma of Pediatric Patients.. Therapeutic drug monitoring. 2023. DOI: 10.1097/ftd.0000000000001128.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.