Berlin Heart EXCOR pediatric ventricular assist device for bridge to heart transplantation in US children.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Almond Christopher S, Morales David L, Blackstone Eugene H, Turrentine Mark W, Imamura Michiaki, Massicotte M Patricia, Jordan Lori C, Devaney Eric J, Ravishankar Chitra, Kanter Kirk R, Holman William, Kroslowitz Robert, Tjossem Christine, Thuita Lucy, Cohen Gordon A, Buchholz Holger, St Louis James D, Nguyen Khanh, Niebler Robert A, Walters Henry L, Reemtsen Brian, Wearden Peter D, Reinhartz Olaf, Guleserian Kristine J, Mitchell Max B, Bleiweis Mark S, Canter Charles E, Humpl Tilman
Аннотация
METHODS AND RESULTS: This multicenter, prospective cohort study involved all children implanted with the Berlin Heart EXCOR Pediatric ventricular assist device at 47 centers from May 2007 through December 2010. Multiphase nonproportional hazards modeling was used to identify risk factors for early (<2 months) and late mortality. Of 204 children supported with the EXCOR, the median duration of support was 40 days (range, 1-435 days). Survival at 12 months was 75%, including 64% who reached transplantation, 6% who recovered, and 5% who were alive on the device. Multivariable analysis identified lower weight, biventricular assist device support, and elevated bilirubin as risk factors for early mortality and bilirubin extremes and renal dysfunction as risk factors for late mortality. Neurological dysfunction occurred in 29% and was the leading cause of death.
CONCLUSIONS: Use of the Berlin Heart EXCOR has risen dramatically over the past decade. The EXCOR has emerged as a new treatment standard in the United States for pediatric bridge to transplantation. Three-quarters of children survived to transplantation or recovery; an important fraction experienced neurological dysfunction. Smaller patient size, renal dysfunction, hepatic dysfunction, and biventricular assist device use were associated with mortality, whereas extracorporeal membrane oxygenation before implantation and congenital heart disease were not.