DOI 10.1016/j.acra.2005.10.004

Медицинская библиотека · Academic radiology · 2006

Cerebral edema attenuated inversion recovery MR sequence in low magnetic field: a feasibility study.

Katisko Jani P A, Yrjänä Sanna K, Tuominen Juho, Kokkonen Salla-Maarit, Ilkko Eero, Erkkilä Jari, Shiffbauer Hagen, Koivukangas John P

Клинический случайСтатьяПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVES

Minimally invasive neurosurgery requires methods to specify surgical boundaries of target tissue, such as brain tumors. This study investigated technical possibilities and clinical usefulness of adapting edema attenuated inversion recovery (EDAIR) pulse sequences to suppress magnetic resonance signal from cerebral edema in brain tumor patients.

MATERIALS AND

METHODS

A resistive 0.23-T magnetic resonance scanner with magnitude-encoded inversion recovery sequences was used. Twenty-eight separate scanning tests in 25 neurosurgical brain tumor patients were performed on the day before surgery. An inversion recovery sequence with several inversion times between 150 and 2,200 ms was tested. The same sequences were also used intraoperatively and postoperatively.

RESULTS

T(1) relaxation time of brain edema varied from case to case. An inversion recovery sequence with an inversion time of 400-800 milliseconds attenuated brain edema and seemed to help in demarcating gross brain tumor for surgical resection. These features were helpful for the evaluation of resectable tumor tissue particularly using neuronavigation techniques.

CONCLUSIONS

According to these preliminary findings, inversion recovery sequences supplement other imaging modalities and assist neurosurgeons in evaluating different surgical trajectories and in estimating brain tumor volume before craniotomy.

Для цитирования: Katisko Jani P A, Yrjänä Sanna K, Tuominen Juho, Kokkonen Salla-Maarit, Ilkko Eero, Erkkilä Jari, Shiffbauer Hagen, Koivukangas John P. Cerebral edema attenuated inversion recovery MR sequence in low magnetic field: a feasibility study.. Academic radiology. 2006. DOI: 10.1016/j.acra.2005.10.004.