Comparison of calcaneal fixation of a retrograde intramedullary nail with a fixed-angle spiral blade versus a fixed-angle screw.
Аннотация
BACKGROUND
Retrograde intramedullary nailing is an established technique for tibiotalocalcaneal arthrodesis (TTCA). In poor bone stock (osteoporosis, neuroarthropathy), device fixation in the hindfoot remains a problem. Fixed-angle spiral-blade fixation of the nail in the calcaneus could be useful.
MATERIALS AND
METHODS
In seven matched pairs of human below-knee specimens, bone mineral density (BMD) was determined, and TTCA was performed with an intramedullary nail (Synthes Hindfoot Arthrodesis Nail HAN Expert Nailing System), using a conventional screw plus a fixed-angle spiral blade versus a conventional screw plus a fixed-angle screw, in the calcaneus. The constructs were subjected to quasi-static loading (dorsiflexion/plantarflexion, varus/valgus, rotation) and to cyclic loading to failure. Parameters studied were construct neutral zone (NZ) and range of motion (ROM), and number of cycles to failure.
RESULTS
With dorsiflexion/plantarflexion loading, the screw-plus-spiral-blade constructs had a significantly smaller ROM in the quasi-static test (p = 0.028) and early in the cyclic test (p = 0.02); differences in the other parameters were not significant. There was a significant correlation between BMD and cycles to failure for the two-screw constructs (r = 0.94; p = 0.002) and for the screw-plus-spiral-blade constructs (r = 0.86; p = 0.014).
CONCLUSION
In TTCA with a HAN Expert Nailing System, the use of a calcaneal spiral blade can further reduce motion within the construct. This may be especially useful in poor bone stock.
CLINICAL RELEVANCE: Results obtained in this study could be used to guide the operating surgeon's TTCA strategy.